Determination of Soil pH
Contents Purpose and Background Extraction Procedure Quality Control References PURPOSE AND BACKGROUNDSoil pH is a measure of hydrogen-ion (H+) activity in a soil suspension. This property influences the many aspects of crop production and soil chemistry, including availability of nutrients and toxic substances, activity and diversity of microbial populations, and activity of certain pesticides. Soil pH is defined as the negative logarithm (base 10) of the H+ activity (moles per liter) in the soil solution. As the activity of H+ in the soil solution increases, the soil pH value decreases. Soils with pH values below 7 are referred to as “acid”; pH values above 7 are referred to as “alkaline”; soils at pH 7 are referred to as “neutral”. EXTRACTION PROCEDURE
QUALITY CONTROL This laboratory participates in the North American Proficiency Testing Program (NAPTA) operated by the Soil Science Society of America for all soil analyses including soil pH. During calibration a 7.0 buffer solution is read by the meter in millivolts. Next a 4.0 buffer solution is read in millivolts (mV). The program calculates a calibration curve using these inputs. A 6.0 buffer is read and after calculation should read exactly 6.0. Four check soils are read and all four-check soils must read within ± 0.1 pH unit. During this calibration there are several tests that have to be passed in order to continue to measure the pH of unknown samples. Two performance tests are used to check electrode performance. In the first performance test the reading of the 7.0 buffer (in mV) added to the reading of the 4.0 buffer (in mV) should equal to at least 172.0 mV. If the value does not meet these criteria then the electrode will have to be recalibrated. If it still does not meet the criteria then a new electrode should be placed on the meter. The second performance test was designed to check the slope of the electrode. The slope was calculated as: Slope = (mV7 – mV4) / (pH7.0 – pH4.0). The slope should be ± 5% of the theoretical Nernst equation slope (59.2 mV at 25°C). If not the electrode is replaced. Every twelfth sample is either a High QC, Low QC, or check soil and must read within ±0.2 of the specified value.
REFERENCESWatson, M.E. and J.R. Brown. 1998. pH and lime requirements. North Central Regional Research Publication. 221: 13-16
Donahue, S.J. 1983. Determination of Soil Water pH. Reference Soil Test Methods for the Southern Region of the United States. p. 1-4. Southern Cooperative Series Bulletin 289. University of Georgia College of Agriculture Experiment Station. Athens, GA.
|